CVE-2017-12718
HIGH 8.1EPSS 12.8%
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
- Affected
- Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
- CVSS v3.0 NVD
- 8.1 HIGH
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H - CVSS v2.0 NVD
- 6.8 MEDIUM
AV:N/AC:M/Au:N/C:P/I:P/A:P - EPSS
- 12.78% chance of exploitation in the next 30 days, 96th percentile
- Published
- 2018-02-15
- Updated
- 2024-08-05
ExploitDB entries (1)
Security advisories
- http://www.securityfocus.com/bid/100665
- http://www.securityfocus.com/bid/101252
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A